Digital insulation resistance meter MIC-5000 is intended for direct measurements of insulation resistance in power cables, transformers, electric motors and other power devices and equipment. Additionally, the device enables measurement of DC and AC voltages as well as low voltage measurement of low resistances. All test terminals of the instrument are fitted with a circuitry protecting the device from damage should the mains voltage appear on the input. The instrument was designed and built in compliance with EN IEC 61557 standard.
- Insulation resistance measurement:
- test voltage selection between 250...5000V in 50V intervals, or predefined 250, 500, 1000, 2500, 5000V,
- on-line insulation resistance measurement or leakage current,
- automatic discharge of tested object’s capacitance after measurement of insulation resistance,
- direct measurement of one or two absorption coefficients,
- acoustic determination of 5 seconds time intervals that facilitates to take time characteristics during insulation resistance measurement,
- saving measured value of test voltages and selected times T1, T2 and T3 ranging from 1... 600 seconds,
- indicator of the actual test voltage value during a measurement,
- protection against live objects. - AC/DC voltage measurement in range 0...600V.
- Memory for 999 measurement results with an ability to transfer the datato a PC.
- Automatic selection of measurement ranges.
- Powered by rechargeable battery pack:
- included with the device an external power supply for automatic battery charging to ensure prolonged battery life and optimal working conditions,
- battery status indicator. - AUTO-OFF function.
- Meter meets the requirements of the standard EN 61557.
Electric security:
- type of insulation: double, according to EN 61010-1 and IEC 61557
- measurement category: CAT III 600V acc. to EN 61010-1
- measurement category for terminals and circuit 5kV: interpolated III 5000V
- protection class acc. to EN 60529: IP54
Rated operational conditions:
- operating temperature: -10...+50°C,
- test voltage accuracy (Robc[Ω]≥1000*Un[V]): -0...+10% of the set value,
- voltage fluctuation due to temperature changes better than: 0,2%/°C,
- converter output current:
min. 1,0mA for rated test voltage
1,0...3,0mA for 2500V
1,0...1,4mA for 5000V
- measurement rate: approx. 1 measure/second
- quality standard: ISO 9001
Other technical data:
- power supply: Ni-MH battery pack
- power supply adaptor: 100...240V
- minimum measurement time 5kV/1mA according to
EN 61557 (5s/25s) 5h or 600 measurements
- time to auto-off:
mode RISO/IL depends on T2 or T3 (T2/T3 + 300 seconds) settings
other modes 300 seconds
Insulation resistance measurement
- test voltage can be set in the range between 250V and 5000V with 50V intervals
- test voltage accuracy (Robc [Ω] ≥ 1000*UN [V]): 0 +10% of the set value
- voltage fluctuation due to temperature changes - better than 0,2% / °C
- measurement times T1, T2 i T3 for absorption coefficient measurement selected from the range between 1sec and 600sec with accuracy ± 1sec
measurement range according to PN-EN 61557-2: RISOmin...5,0TΩ; RISOmin=UISOnom/1mA
Range | Resolution | Accuracy |
0,0...999,9kΩ | 0,1kΩ | ±(3% m.v. + 20 digits) |
1,000...9,999MΩ | 0,001MΩ | |
10,00...99,99MΩ | 0,01MΩ | |
100,0...999,9MΩ | 0,1MΩ | |
1,000...9,999GΩ | 0,001GΩ | |
10,00...99,99GΩ | 0,01GΩ | |
100,0...999,9GΩ | 0,1GΩ | |
1,000...5,000TΩ | 0,001TΩ |
• for measurements of limited converter current, the accuracy is not specified.
Minimum insulation resistance measured without limited converter current
Test voltage | Range |
250V | 250kΩ |
500V | 500kΩ |
1000V | 1,0MΩ |
2500V | 2,5MΩ |
5000V | 5,0MΩ |
DC voltage measurement
Range | Resolution | Accuracy |
0...600V | 1V | ±(3% m.v. + 2 digits) |
AC voltage measurement
50Hz (sinusoidal shape with harmonic contents < 2%)
Range | Resolution | Accuracy |
0...600V | 1V | ±(3% m.v. + 2 digits) |
„m.v.” - measured value.
Leakage current reading
Range | Resolution | Accuracy |
0...Ipmax | depending on range |
-ΔI-, +ΔI+ |
Ipmax - maximum converter current 1,2 ± 0,2 mA,
ΔI-, +ΔI+ - current reading accuracy calculated using the following formula:
UISO - test voltage
R - displayed insulation resistance value
ΔR - resistance measurement defined for a particular measurement